Visual troubleshooting library

Acquisition, caliper placement, waveform selection, common errors and artifacts. Items without a cleared image use an honest description rather than a placeholder clip.

Native valves

Visual troubleshooting library

Left ventricle

  • Foreshortening — Apical four- and two-chamber views at reduced depth focused on the LV cavity; dedicated atrial views separately.
  • Strain tracking — Three apical views at reduced sector width including only the ventricles, true apex visualized, speckle tracking of the full wall.
  • Regional wall thickening — Parasternal short axis at the papillary-muscle level (circular ventricle, papillary muscles not wobbling) plus apical views with the sector reduced to the ventricles; each segment in more than one view.
  • Apical densities — Apical views with the focus and depth set for the near field; PSAX at the apical third (transducer down an interspace and lateral); colour Doppler with a low scale; UEA when uncertain.
  • LV internal dimension: caliper placement — Parasternal long axis along the centre axis of the ventricle (papillary muscles not visible); biplane imaging can confirm perpendicular alignment.

Atria and diastolic function

Right heart and pulmonary pressures

  • RV-focused acquisition — RV-focused apical four-chamber: from the A4C, rotate slightly counterclockwise at the apex to maximize RV area, keeping the LV apex centred and avoiding anterior tilt into a five-chamber view.
  • Septal motion — PSAX at the papillary-muscle level for septal position and LVEI; A4C observed during respiration for interdependence; M-mode for timing.
  • IVC measurement site — Subcostal long axis showing the IVC for several centimetres as it enters the right atrium, with the hepatic vein; patient supine.
  • TR peak velocity signal selection — CW Doppler of the TR jet from multiple windows (apical, parasternal RV inflow, subcostal), colour-guided to align with the jet.