Skip to content
EurekaE/e'ureka: Echo Reading
Text size 100%

Aliasing and range ambiguity

ViewPulsed-wave and colour Doppler; CW and HPRF as alternatives.
TimingPeaks are cut off and wrap to the opposite side of the baseline when velocity exceeds the Nyquist limit.
Caliper / traceShift the baseline, lower the transducer frequency, reduce sample depth (raising PRF), or switch to CW/HPRF accepting range ambiguity.
No image yet
No cleared image or clip is available for this item yet; the description below stands in. Fabricated or stock images are not used.

Correct example

An LVOT PW signal that wraps in diastole because of aortic regurgitation is re-recorded with CW, which displays the full AR velocity above the baseline; the LVOT systolic velocity itself is kept on PW with baseline shift.

Common error

Measuring the peak of a wrapped signal, mis-reading its direction, or assuming an HPRF or CW velocity came from the sample-volume depth.

Effect on interpretation

Aliasing precludes accurate peak velocity and confuses direction; CW and HPRF remove aliasing but cannot locate the depth of the peak, so the origin of a high velocity must be inferred from imaging and colour.

Why

The Nyquist limit is half the PRF, and PRF falls with sample depth; a non-coaxial angle separately underestimates velocity without any visible sign, which is why stenotic jets are interrogated from multiple windows.