Tissue Doppler e′ sampling (septal and lateral annulus)

ViewApical four-chamber, pulsed tissue Doppler at the septal (medial) and lateral mitral annulus, beam as parallel as possible to annular motion.
TimingEarly diastolic annular velocity (e′); sweep speed 100 mm/s.
Caliper / traceLarge 5–10 mm sample volume at the annulus; scale at or below about 25 cm/s; peak e′ measured; lateral and medial values averaged for E/e′ and also reported individually.
No image yet
No cleared image or clip is available for this item yet; the description below stands in. Fabricated or stock images are not used.

Correct example

A system-specific DTI preset, gate covering annular motion, scale reduced to enlarge the signal, and both annular sites recorded so the averaged e′ can be used.

Common error

Using a conventional PW preset, a small gate that drops out of the annulus, an angulated beam, or only one annular site when the average is required.

Effect on interpretation

Absolute e′ and the E/e′ ratio are major parts of the non-invasive filling-pressure criteria, so sampling errors move patients across grading cut-offs.

Why

Tissue Doppler measures low-velocity, high-amplitude wall motion, which is why the filter, scale and gate differ from blood-flow Doppler.